MONOLITH Gamma & X-ray HPGe Spectrometer based on Pulse-tube
Description
Spectrometers based on HPGe (high purity germanium) detectors are designed to detect gamma radiation in the energy range from 40 keV to 3 MeV for GCD models and from 3 keV to 3 MeV for GCDX models.
Brochure Monolith Gamma- ray HPGe Detector Electrically cooled
Download | PDF
Application
Registration, accumulation and processing of X-ray and gamma radiation spectra under conditions when the use of liquid nitrogen for cooling HPGe detectors is undesirable or impossible.
Features
- 10% – 160% efficiency HPGe p-type coaxial detectors are available
- Low background and Ultra – low background materials are available
- Manufacture in a portable cryostat is possible
Precision
- Adopting precision gamma-spectrometry methods
- High efficiency of radiation detection
- Excellent peak symmetry
Reliability
- Energy range from 40 keV to 10 MeV for GCD model
- Energy range from 3 keV to 10 MeV for GCDX/GCDX-OS models
- High energy rate up to 200000 MeV/sec
Specification
Relative efficiency (with respect to 3’’ x 3’’ NaI detector and Co-60 source mounted 25 cm above the detector) at 1.33 MeV g-photon:
30 %
Peak to Compton ratio:
52 : 1
Energy resolution
at 122 keV: < 1000 eV
at 1.33 MeV: < 2000 eV
Energy range of detector operation:
3 keV – 3 MeV
Peak shape
FWTM/FWHM: < 1.9
FW.02M/FWHM: < 2.65
Endcap window material:
Al / Be / Carbon fibre
Cooling time of the detector, hours
< 13
Autonomous operation time, hours:
24/7
Power:
230 V, 50/60 Hz
Power consumption, max:
250 W
Power consumption, normal:
120 W
Range of the operating temperatures, ° С:
-10 … +40
30 %
Peak to Compton ratio:
52 : 1
Energy resolution
at 122 keV: < 1000 eV
at 1.33 MeV: < 2000 eV
Energy range of detector operation:
3 keV – 3 MeV
Peak shape
FWTM/FWHM: < 1.9
FW.02M/FWHM: < 2.65
Endcap window material:
Al / Be / Carbon fibre
Cooling time of the detector, hours
< 13
Autonomous operation time, hours:
24/7
Power:
230 V, 50/60 Hz
Power consumption, max:
250 W
Power consumption, normal:
120 W
Range of the operating temperatures, ° С:
-10 … +40
Complete set
- HPGe detector
- Preamplifier
- Autonomous cooling system for the detector based on electrical machinery cooler EMC
- Controller for controlling the operation of EMC
- Fans (2-4) for EMC cooling
Options
- Software package NuclideMasterPlus
- Advanced software package
- Multichannel Analyzer