TIM-602 Gamma Activity Monitor
Description
High sensitive activity monitor based on a scintillation crystal (NaI(Tl), 63×63 mm.), designed to measure specific activity of 137Cs of different substances
Brochure TIM-602 Gamma Activity Monitor
Download | PDF
Sertificate
Download | PDF
GammaSP_v1.0.6
Download | ZIP
Application
- Radiation monitoring of drinking water
- Radiation monitoring of food and raw materials
- Monitoring of radioactive contamination of industrial products and industrial waste
- Radiation monitoring of forestry products
Features
General
- Hight sensitivity
- Large crystal, NaI(Tl), ø63х63 mm.
- Spectra processing and calculation values by “energy windows” method
- Easy to use
Reliability
- Spectrum stabilization by 40K
- Check of energy calibration
- Methodological support of measurements
Additional functions
- Promt check for compliance with a control level
- Store and presentation of gamma spectrum of current measurement
- GammaSP software package installed on a PC for operating the monitor
Specification
Detector type:
NaI(Tl), ø63х63 mm.
Measurement range of specific activity 137Cs:
5 – 10000 Bq/kg
Limit of activity measurement intrinsic relative error:
±20%
Density range of controlled substances:
0.1 – 2.0 kg/dm3
Sample volume:
0.1; 0.3; 0.5 and 1 l.
Energy range:
50 – 3000 keV
Energy resolution (FWHM) at 137Cs (662 keV):
8.5 – 11%
Number of ADC channels:
1024
Setting time:
5 min.
Temperature range:
-10 – 40°C
Supply voltage:
230 ± 10% V
Power consumption:
20 ВА
Lead shield wall thickness:
50 mm
Weight:
Detector: 2 kg
Lead shield: 120 kg
Interface:
Русский
English
Беларуская
NaI(Tl), ø63х63 mm.
Measurement range of specific activity 137Cs:
5 – 10000 Bq/kg
Limit of activity measurement intrinsic relative error:
±20%
Density range of controlled substances:
0.1 – 2.0 kg/dm3
Sample volume:
0.1; 0.3; 0.5 and 1 l.
Energy range:
50 – 3000 keV
Energy resolution (FWHM) at 137Cs (662 keV):
8.5 – 11%
Number of ADC channels:
1024
Setting time:
5 min.
Temperature range:
-10 – 40°C
Supply voltage:
230 ± 10% V
Power consumption:
20 ВА
Lead shield wall thickness:
50 mm
Weight:
Detector: 2 kg
Lead shield: 120 kg
Interface:
Русский
English
Беларуская
Complete set
- Detection unit
- Processing unit
- Lead shield
- Control source based on 40K
- Marinelli beaker 1 l.
- Flat wessel 0,3 l.
- Operational documentation
- Methodological documentation